3 - 62
The coefficients of internal DH200 transducers used in the Model 9016 are
selected with array indexes aa=01 through 10 (hex). Coefficients of external
9400 transducers used by Models 9021 and 9022 (digitally compensated) are
selected with array indexes aa=01 through 0C (hex). All valid coefficient indexes
(for each of these arrays) are listed in the following table:
cc
Transducer Coefficient Description
Datum Type
00
Re-zero Cal Adjustment (offset) term
FLOAT
01
Span Cal Adjustment (gain) term
FLOAT
02
Dynamic EU Conversion coefficient c0
FLOAT
03
Dynamic EU Conversion coefficient c1
FLOAT
04
Dynamic EU Conversion coefficient c2
FLOAT
05
Dynamic EU Conversion coefficient c3
FLOAT
06
Reserved for Factory Use
----
07
User Defined Date field (see end-of-table note)
INTEGER
08
Date of Factory Calibration (see end-of-table note)
INTEGER
09
Transducer Manufacturing Reference number
INTEGER
0A
Transducer Full-Scale Range code (see Appendix
F)
INTEGER
0B-0F
Temperature 1, Pressures 1-5 voltages
FLOAT
10-14
Temperature 2, Pressures 1-5 voltages
FLOAT
15-19
Temperature 3, Pressures 1-5 voltages
FLOAT
1A-1E
Temperature 4, Pressures 1-5 voltages
FLOAT
1F-23
Temperature 5, Pressures 1-5 voltages
FLOAT