Measurement Specialties Inc. NetScanner System (9016, 9021, & 9022) User’s Manual
4 - 13
4.5 9021/9022 Analog Calibration
Both the standard 9021 and the 9022 modules are used to make absolute analog measurements
from either the PSI 9400 transducers or third party analog output transducers. In order to maintain
the highest level of analog accuracy, the 9021/9022 is designed to allow its firmware to perform
automatic software calibration of the analog interface circuitry. This automatic, firmware
controlled calibration function minimizes analog offset and gain errors for each of the twelve
9021/9022 input channels. This auto-calibration function reduces analog measurement errors
resulting from long term or thermally induced component drift. Unless disabled through
software commands to the 9021/9022, this software calibration function will be constantly
performed by the 9021/9022 firmware, therefore maintaining consistent measurement
accuracy within the 9021/9022. In order to perform this software calibration function, the
9021/9022 contains one internal voltage reference network. This internal calibration standard is
designed to provide excellent long term and thermal stability. This use of a common reference
network reduces user or factory calibration of the 9021/9022 to a simple calibration of this one
internal reference network. The following procedures describe the methods for verifying analog
measurement accuracy and for calibrating this internal reference in order to maintain optimum
accuracy through the 9021/9022.
NOTE: When using the 9021/9022 with digitally compensated PSI Model 9400
transducers, the span calibration functions (for the 9400 transducers) can serve as a
method to correct both 9400 and internal 9021/9022 measurement accuracies.
Periodic use of this span calibration function can eliminate the need for the described
voltage reference calibration when using the 9021/9022 with digitally compensated
Model 9400 transducers.