Measurement Specialties Inc. NetScanner System (9016, 9021, & 9022) User’s Manual
4 - 11
as it does for Re-zero adjustment, it should be placed in the desired position manually with the
Set Operating Options (‘w’) command (option indexes = 0C and 12). This is illustrated in the
example of the next section.
When multi-point calibrating Model 9016 modules with multiple ranges installed, the CAL port
may be used to apply pressure to all transducers only if the specified proof pressure is not
exceeded on any channel. If the application of a specific span pressure exceeds the proof
pressure rating of any other transducer contained within the same scanner, the calibration
pressures must be applied to the RUN side pneumatic input ports. Since the calibration
command (‘C’) has a channel selection bit map parameter allowing it to calibrate only the desired
pressure channels, the RUN port is a viable option for supplying the calibration pressures.
4.4.2 Multi-Point Calibration Summary
Following is a simple step-by-step procedure for executing a “multi-point” calibration of a 9016,
9021, or 9022 Intelligent Pressure Scanner. For Models 9021 and 9022, skip the steps that
manipulate valves as denoted by the double asterisk (**). It is assumed that all channels in the
unit have the same full-scale pressure range. Optional commands are shown within brackets [ ].
Should it become necessary to abandon this calibration procedure once it is started, you may
execute the Abort sub-command [‘C 03'] of ‘C’ at any time after the first ‘C’ sub-command.
Description
9016
Command
9021/9022
Command
Ensure that valves in RUN/CAL mode (default).
[w1200]
**
… normal data acquisition assumed to be running
Place the module calibration manifolds into the CAL
position if this is the desired pressure application input.
The RUN position may be a better choice for modules
with transducers having different ranges.
[w0C01]
for CAL
[w0C00]
for RUN
**
Ready the module for multi-point calibration by
executing the Configure & Start (‘00’) sub-command of
‘C’. This establishes all the channels to be affected, and
determines the total number of calibration points that will
be supplied (3 in this example) in later steps. It also
starts module averaging for calibration (64 samples in this
example). The linear fit (1) is required.
C 00 FFFF 3 1 64 (9016)
or
C 00 0FFF 3 1 64 (9021/9022)