Measurement Specialties Inc. NetScanner System (9016, 9021, & 9022) User’s Manual
4 - 5
4.2.2 Re-zero Calibration Summary
Following is a simple, step-by-step procedure for executing a Re-zero calibration of a 9016 or
9021/9022 Intelligent Pressure Scanner. For Models 9021 and 9022, skip the steps that
manipulate valves as denoted by a double asterisk (**). Optional commands are shown within
brackets [ ].
Description
9016
Command
9021/9022
Command
Disable automatic valve shifting after module power up. Ensure
valves in RUN/CAL mode (default)
[w0B01]
[w1200]
**
normal data acquisition
Apply 0.0 psi differential to the module CAL and CAL REF
inputs.
Place the module calibration manifolds into the CAL position [if
w0B01 command executed in Step 1]
w0C01
**
Delay for settling of pneumatic inputs
Verify that measured data reads near expected zero value
[rFFFF0]
[r0FFF0]
Instruct module to calculate new offset coefficients for all 16
channels of Model 9016 (all 12 for Models 9021and 9022)
hFFFF
h0FFF
Place calibration manifold back into the RUN
position [if w0B01
command executed in Step 1]
[w0C00]
**
Store new offset coefficients to transducer nonvolatile memory
[w08]
continue normal data acquisition