Pressure Systems 9022 Instrukcja Użytkownika Strona 116

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 115
www.meas-spec
.com
Measurement Specialties Inc. NetScanner System (9016, 9021, & 9022) User’s Manual
4 - 4
4.2 Re-zero Calibration
The NetScanner
System Intelligent Pressure Scanners including Models 9016, 9021, and 9022
are capable of independently performing a transducer calibration adjustment function, referred to
as Re-zero adjustment (or simply Re-zero Cal). The Re-zero calibration will result in a
recalculated “offset” coefficient for each channel being calibrated, which automatically
compensates for any transducer offset drift errors. Since the factory-set coefficients in each
transducer (that characterize both pressure and temperature) are extremely stable over time, these
simple offset corrections compensate for the majority of transducer errors over time. For this
reason, a Re-zero may be the only calibration adjustment required by many applications. For
those applications requiring more accuracy, an optional single-point Span-only calibration
adjustment will be described in Section 4.3. An improved multi-point calibration adjustment,
integrating both the Re-zero and Span calibration process will be described in Section 4.4.
For Model 9016, with integral DH200 pneumatic transducers, internal manifolds and valves
allow a Re-zero calibration to be accomplished easily and automatically. For Models 9021 and
9022, which have external 9400, 9401 or 9402 (or third-party) all-media transducers, the user
must supply any necessary valves and controls to accomplish the application of a “minimum”
pneumatic or hydraulic pressure to these transducers before executing the re-zero adjustment.
When instructed to execute a Calculate and Set Offsets (‘h’) command, the NetScanner
System module will automatically perform the Re-zero adjustment calibration, and then update
the offset coefficients in its volatile memory. It will subsequently use the newly calculated terms
for all future engineering-unit calculations until power is lost to the module.
4.2.1. Re-zero Calibration Valve Control
When instructed to execute a Re-zero (Calculate and Set Offsets (‘h’)) command, Model 9016
modules will normally shift the internal calibration valve into the CAL position and use the
pressures present at the CAL and CAL REF inputs for the “minimum” (e.g., zero) calibration
pressure. After the Re-zero adjustment is complete, the Model 9016 calibration valve will be
placed in the RUN position. Models 9021 and 9022 assume a “minimum” calibration is already
applied to the transducers. This automatic shift of the calibration valve can be disabled through
use of the Set Operating Options (‘w’) (option index=0B hex) command. This allows
independent control of the calibration valve by the user using other options (see option indexes =
0C and 12 hex) of the same command.
NOTE: When using the Calculate and Set Offsets command (‘h’), only local terms in the
module’s volatile main memory (RAM) are updated. Under normal operation, it is not
recommended to store these new coefficients in transducer non-volatile memory.
Instead, the re-zero should be performed at regular intervals. Refer to Section 4.6 if it is
desired to also store these new re-zero terms in transducer nonvolatile memory.
Przeglądanie stron 115
1 2 ... 111 112 113 114 115 116 117 118 119 120 121 ... 193 194

Komentarze do niniejszej Instrukcji

Brak uwag